Search JIC Articles:
The Official Journal of the International Andreas Gruentzig Society
Saturday, September 6, 2008


text size: A A A
manuscripts: submit | review
A Novel Use for a Distal Embolic Protection Device: Stent Retrieval
CASE REPORTS:
A Novel Use for a Distal Embolic Protection Device: Stent Retrieval

- Pavel Guigauri, MD and Harold L. Dauerman, MD


       Stent embolization is an uncommon but potentially serious complication of percutaneous coronary intervention (PCI). The incidence of coronary stent embolization is 0.3–8.3% and varies with stent type, need for manual crimping, and clinical situation.1–4 There are numerous case reports describing methods for retrieving embolized coronary stents. These methods include the use of angioplasty balloons and snares.5–9 In some of these situations, the goal has been complete removal of the stent. In others, the goal has been simply to retrieve the stent from the coronary circulation in favor of release of the stent into the peripheral circulation. Despite these methods, the stent retrieval is unsuccessful in at least 30% of these cases and serious consequences may result.3,4 Given
Figure 1a
(A) The saphenous vein graft to the obtuse marginal branch has 2 severe lesions (white arrows).
Figure 1b
(B) The Filterwire is placed in the distal portion of the graft. Still, the proximal edge of the Filterwire is too close to the distal vein graft lesion to allow placement of an Express-2 stent covering the entire lesion.
the difficulty in utilizing snares and balloons for stent retrieval, we describe a new method of stent retrieval utilizing a distal embolic protection device.

       Case Report. A 61-year-old male with a history of coronary bypass grafting ten years prior to admission was referred for cardiac catheterization. The indication for cardiac catheterization was unstable angina. Angiography revealed a 90% stenosis in the ostium of the SVG to the obtuse marginal branch, as well as an 80% lesion in the distal portion of the graft (Figure 1A). Stenting with distal embolic protection was planned, utilizing a 7 Fr Amplatz guide and a Filterwire (Boston Scientific, Natick, Massachusetts).
       After positioning the Filterwire in the most distal portion of the vein graft (Figure 1B), an Express2 4.5 x 12 mm stent (Boston Scientific) was delivered to the distal graft, but the lesion could not be covered due to the Filterwire location. Thus, an attempt was made to withdraw the stent into the guide and we considered other options for distal SVG stenting. As the stent entered the Amplatz guide, it was stripped off the delivery balloon, leaving it free on the proximal portion of the Filterwire (Figure 1C). The stent could not be retrieved with an Amplatz Goose Neck® Snare (ev3, Plymouth, Minnesota). A buddy wire was placed in order to attempt passage of the snare over the buddy wire or passage of a second balloon distal to the stent (with the hope that low pressure inflation of the distal balloon would allow the stent to be trapped proximal to the balloon and pulled back into the guide). Neither of these techniques were successful. The stent was retrieved using a novel technique (Figure 1D): The fully deployed filterwire was pulled into the guide with the stent trapped and crushed into the tip of the guide. While maintaining traction on the wire-guide combination, the entire guidewire-stent assemblywas removed intact via the femoral artery sheath. Subsequently, the graft lesions were wired without distal embolic protection and stents were
Figure 1c
(C) The Express stent is free on the Filterwire after the attempted retrieval of the stent into the Amplatz guide (white arrow). A second buddy wire has been placed to attempt snare and second balloon retrieval methods (white arrow). Also, the Filterwire has slipped proximal to the distal lesion during these manipulations (black arrow).
Figure 1d
(D) The Filterwire is withdrawn and the stent is crushed into the guiding catheter (white arrow) allowing withdrawal of stent, wire and guide into the femoral sheath.
placed proximally and distally without complications.


       Discussion. The largest stent embolization registry reported to date included 387 patients with 411 stent embolizations.4 The technical challenge afforded by stent embolization is emphasized in this study; successful stent retrieval was achieved in only 30% of the cohort, and mortality among patients with unsuccessfully retrieved stents was 17% versus 0.9% for those with successful stent retrieval. While other studies have had a higher success rate for retrieval of embolized stents,3 there is at least a minority of patients in whom embolized stents will truly be “lost” in the distal vasculature due to the inability of snare and balloon-based techniques to capture the embolized stent. Given the technical challenges and serious consequences associated with coronary stent embolization, new techniques to improve stent retrieval are of interest. We present here a novel method for retrieving a coronary stent utilizing a distal embolic protection device.
       The Filterwire distal embolic protection device is one of two FDA-approved techniques for improving saphenous vein graft stenting.10 The construction of this particular distal embolic protection device (basket retrieval) makes it theoretically not only a capture device for emboli, but for larger objects as well. In this particular case, it was utilized to capture a stent lost on the proximal wire. Note was made that this technique was only utilized after
Figure 1e
(E) Final result after rewiring and stenting the vein graft without use of distal protection device.
standard techniques failed (second wire with attempted distal balloon inflation, and attempted retrieval with a snare). Also, notably, retrieval of the fully expanded Filterwire basket through two severe vein graft lesions did not lead to acute disruption of the vein graft. Our case study is consistent with a prior report describing the similar utilization of distal balloon occlusion device for stent retrieval.11 The current case study extends these observations to the utilization of a basket retrieval-type distal protection device.
       The incidence of stent embolization has clearly decreased since the era of manual crimping of coronary stents.1,4 Nevertheless, the rare occurrence of this PCI-related complication can present a challenging technical situation for interventional cardiologists. Given the infrequency of embolization, interventional cardiologists may be expected to be less well trained in the use of snare techniques, making retrieval rates even lower than the 30–70% rates seen in prior studies.3,4 The use of basket retrieval devices for stent embolization is an attractive option. The use of this option in other situations (i.e., when the stent is not free on the already deployed Filterwire) may be feasible; as lower-profile basket retrieval devices are developed, the ability to pass the embolized stent without pushing the free stent forward will determine the applicability of such an approach.


1. Eggebrecht H, Haude M, von Birgelen C, et al. Nonsurgical retrieval of embolized coronary stents. Cathet Cardiovasc Intervent 2000;51:432–440.
2. Alfonso F, Martinez D, Hernandez R, et al. Stent embolization during intracoronary stenting. Am J Cardiol 1996;78:833–835.
3. Dunning DW, Kahn JK, O'Neill WW. The long-term consequences of lost intracoronary stents. J Intervent Cardiol 2002;15:345–348.
4. Bolte J, Neumann U, Pfafferott C, et al. Incidence, management, and outcome of stent loss during intracoronary stenting. Am J Cardiol 2001;88:565–567.
5. Kobayashi Y, Nonogi H, Miyazaki S, et al. Successful retrieval of unexpanded Palmaz-Schatz stent from left main coronary artery. Cathet Cardiovasc Diagn 1996;38:402–404.
6. Pan M, Medina A, Romero M, et al. Peripheral stent recovery after failed intracoronary delivery. Cathet Cardiovasc Diagn 1992;27:230–233.
7. Berder V, Bedossa M, Gras D, et al. Retrieval of a lost coronary stent from the descending aorta using a PTCA balloon and biopsy forceps. Cathet Cardiovasc Diagn 1993;28:351–353.
8. Douard H, Besse P, Broustet JP. Successful retrieval of a lost coronary stent from the descending aorta using a loop basket intravascular retriever set. Cathet Cardiovasc Diagn 1998;44:224–226.
9. Cishek MB, Laslett L, Gershony G. Balloon catheter retrieval of dislodged coronary artery stents: A novel technique. Cathet Cardiovasc Diagn 1995;34:350–352.
10. Baim DS, Wahr D, George B, et al. Randomized trial of a distal embolic protection device during percutaneous intervention of saphenous vein aorto-coronary bypass grafts. Circulation 2002;105:1285–1290.
11. Webb JG, Solankhi N, Carere RG. Facilitation of stent retention and retrieval with an emboli containment device. Cathet Cardiovasc Intervent 2000;50:215–217.

The Journal of Invasive Cardiology - ISSN: 1042-3931 - Volume 17 - Issue 03 (March 2005) - March 2005 - Pages: 183 - 184



Novel Approaches to Managing Bradycardia during Coronary Rheolytic Thrombectomy

Special Supplement to the Journal of Invasive Cardiology


This special supplement was made possible through a grant from Possis Medical, Inc.
CLINICAL EXPERIENCE WITH A NEW HYBRID CORONARY WIRE

On Demand Web Archive
Non-Accredited


This activity is supported by an educational grant from Terumo Medical Corporation.
Pharmacoinvasive Management of Acute Coronary Syndrome: Incorporating the 2007 ACC/AHA Guidelines

Complimentary Accredited CME Program

This activity is supported by an educational grant from Sanofi-Aventis.
Varicose Veins: Causes, Symptoms, Diagnosis and Treatment of Chronic Venous Insufficiency

A Complimentary Accredited ON-DEMAND Webcast

This activity is supported by an educational grant from Diomed, Inc.


Create a Successful Vena Cava Filter Practice

Accredited CD

This activity is supported by an educational grant from Cook Incorporated and has been designed for Interventional Cardiologists, Vascular Surgeons, Fellows and Interventional Cardiovascular Nurses and Technologists.

Achieving Optimal Outcomes in Carotid Stenting: Lessons Learned from Recent Clinical Trials
Complimentary Accredited ON DEMAND Webcast

Topics
1. EVA-3S & Space-Bumps in the road
2. CAPTURE 3500-Lesion morphology & Predictors for Stroke
3. CAPTURE II vs. EXACT 1500-Does open or Closed Cell Stent design really matter?

This activity has been developed for Interventional Cardiologists, Vascular Surgeons, Interventional Radiologists, Neurologists, Interventional Nurses and Technologists with an interest in the diagnosis and treatment of peripheral artery disease.

Anticoagulation Techniques for Peripheral Vascular Interventions

Complimentary Accredited ON DEMAND Webcast

This activity has been developed for Interventional Cardiologists, Vascular Surgeons, Interventional Radiologists, Podiatric Physicians, Endovascular Allied Professionals, Endocrinologists, Wound Care Specialists, Directors of the Wound Care Clinic, and Primary Care Physicians, Pharmacists, Nurses and Technologists.


March 2007 Supplement
On-Demand Webcast
Archived Webcast

About HMP Communications

HMP Communications LLC (HMP) is the authoritative source for comprehensive information and education servicing healthcare professionals. HMP’s products include peer-reviewed and non-peer-reviewed medical journals, national tradeshows and conferences, online programs and customized clinical programs. HMP is a wholly owned subsidiary of HMP Communications Holdings LLC, which also owns the North American Center for Continuing Medical Education (NACCME). NACCME provides a wide array of accredited CME offerings with industry thought leaders participating in roundtable meetings, webcasts, symposia, conferences, seminars, podcasts and satellite programs. Discover more about HMP’s products and services at www.hmpcommunications.com. ©2008 HMP Communications


© 2008 HMP Communications | All Rights Reserved
83 General Warren Blvd, Suite 100 | Malvern, PA 19355 | 800.237.7285