Tissue Shearing with Distal Translocation Complicating Cutting Balloon Angioplasty During Treatment of In-stent Restenosis
CLINICAL DECISION MAKING:
Tissue Shearing with Distal Translocation Complicating Cutting Balloon Angioplasty During Treatment of In-stent Restenosis

- Christopher H. Smith, Michael B. Erwin, Efthymios N. Deliargyris


       Case report. A 49-year-old white female presented with
Figure 1
Figure 1B
recurrent angina. Six months prior, she was successfully treated for a similar anginal syndrome with a 5.0 x 43 Magic Wallstent in the proximal and mid-RCA. Coronary angiography showed a focal 90% narrowing in the proximal segment of the Magic Wallstent; no other lesions were noted (Figure 1A). Due to the focal character of the restenotic lesion, angioplasty as primary treatment was considered acceptable and intracoronary brachytherapy was not contemplated.5 The patient was treated with 3,500 units of heparin and weight-based abciximab bolus and infusion. The lesion was treated with a 5.0 x 18 mm High-Sail balloon to 4 atm, but significant residual narrowing was present. The lesion was then
Figure 1C
Figure 1D
treated with a 4.0 x 15 mm cutting balloon inflated up to 11 atm. Following treatment with the cutting balloon, a filling defect was noted distal to the site of the balloon inflation, but still within the stented segment (Figure 1B). Activated clotting time was 273 seconds. Intracoronary ultrasound (Endodonics) showed a hypoechogenic mass more consistent with tissue rather than thrombus. Repeat dilatation was attempted with a 5.0 x 22 mm NC Big Raptor to 6 atm, however, the severity and location of the lesion was unchanged. An attempt at tissue removal utilizing a 2.33 mm TEC catheter was unsuccessful (Figure 1C). A Multilink Ultra 5.0 x 28 mm stent was positioned covering the mass and deployed inside the existing stent at 13 atm (Figure 1D). The narrowing from the mass was still not completely
Figure 1E
Coronary angiogram of the right coronary artery in a left anterior oblique projection with minimal cranial angulation. (A) Focal severe narrowing within the proximal segment and mild-to-moderate narrowing in the middle of the 5.0 x 43 mm Magic Wallstent placed in the RCA six months earlier. (B) Angiogram following treatment of the proximal narrowing with CBA. Note a good result in the proximal segment, but the appearance of a new, severe narrowing in the middle part of the stent. (C) Attempts to remove the lesion with a 2.33 mm TEC catheter. (D) Deployment of a 5.0 x 28 mm Ultra Multilink stent over the lesion and within the existing Wallstent (“stent sandwich”). (E) Final angiogram demonstrating a mild residual narrowing within the “stent sandwich” despite post-dilation up to 18 mm.
relieved and the stent was further expanded to 18 atm with the 5.0 x 18 mm High-Sail balloon, but the mass proved resistant and a residual 15% narrowing was still evident (Figure 1E). The patient recovered well and reported resolution of her angina.

References (see below)


How Would You Treat This Patient?

David R. Ramsdale, MD
The Cardiothoracic Centre
Liverpool, United Kingdom


       This case deals with the management of focal in-stent restenosis in a large right coronary artery previously stented with a 5.0 x 43 mm Magic Wallstent. Neointimal tissue that causes in-stent restenosis is often relatively soft and responds well to balloon angioplasty. I would agree with the authors that the best treatment would be to use balloon angioplasty. If the lesion is resistant and the angioplasty results are suboptimal, then cutting balloon angioplasty could be used next. Some operators prefer to use cutting balloon as the initial treatment in focal cases of in-stent restensosis, as this is thought to be more effective in disrupting the neointimal hyperplasia that causes in-stent restenosis. However, there is a possibility of damaging the stent struts with the blades of the cutting balloon, thus, I would reserve its use for resistant lesions.
       The operators noticed a new filling defect distal to the site of cutting balloon inflation which could be due to thrombus or tissue displacement. Thrombus formation within the stent is unlikely with a therapeutic activated clotting time and good distal flow. The fact that the lesion proved resistant to further balloon angioplasty confirmed that it was tissue. Placing a stent within the previous stent is often used in such cases to try to compress the tissue between the two layers of metal. Due to the large size of the vessel, drug-eluting stents are not an option, as they are only available up to a maximum nominal diameter of 3.5 mm. A Multilink Ultra 5.0 x 28 mm with a high metal-to-artery ratio was deployed within the previous stent, with a 15% residual narrowing.

Nicolas W. Shammas, MS, MD, FSCAI
Midwest Cardiovascular Research Foundation
Genesis Heart Institute Affiliate
Cardiovascular Medicine, PC
Davenport, Iowa


       Although the culprit segment of the lesion illustrated in this case appears to be focal at the proximal edge of the Magic Wallstent, the proximal and mid-portion of the stent also shows a lesser degree of restenosis. In our experience, we find it that treating only a short segment of a restenotic lesion, without addressing the remaining restenotic segments of that same lesion (though less severe), quite often leads to suboptimal angiographic results. My approach to this lesion will be to initially use a long 4.0 x 30 mm FX Minirail balloon (Guidant) and to dilate the entire restenotic segment at the same time. The advantages of this approach are: 1) less “watermelon seeding” with the Minirail as compared to the use of a regular balloon; 2) an adequate coverage of the entire restenotic lesion is likely to be achieved; 3) the 4.0 mm balloon at 10 to 12 atmospheres leads to a theoretical minimal luminal diameter of 4.5 to 4.57 mm, which should be adequate for this large vessel size; and 4) coverage for the entire lesion will likely avoid tissue shifting and prolapse. If results remain suboptimal despite this approach, (proximal dissection, plaque shifting, continued significant residual stenosis or intraluminal tissue prolapse), then I will use adjunctive stenting. Generally, restenting the vessel leads to optimal angiographic results as in this type of case. If tissue shifting was the reason for adjunctive stenting, it is important to have a long stent to adequately cover several millimeters proximal and distal to the affected area to avoid further shifting of the tissue beyond the restented segment. Drug-eluting stents (DES) in this case are not ideal, given the large vessel size. The current 3.5 mm DES on the market in the United States can be expanded to 4.0 mm. Larger expansion might lead to polymer delamination and loss of effectiveness of the stent. Larger DES stents might become available in the near future.


1. Moreno PR, Palacios IF, Leon MN, et al. Histopathologic comparison of human coronary in-stent and post-balloon angioplasty restenotic tissue. Am J Cardiol 1999;84:462–466.
2. Adamian M, Colombo A, Briguori C, et al. Cutting balloon angioplasty for the treatment of in-stent restenosis: A matched comparison with rotational athrectomy, additional stent implantation, and balloon angioplasty. J Am Coll Cardiol 2001;38:672–679.
3. Huang JW, Hsien LK, Chiau SL, Yuan TL. Coronary stent strut avulsion in aorto-ostial in-stent restenosis: Potential complication after cutting balloon angioplasty. Cathet Cardiovasc Intervent 2002;56:215–219.
4. Haridas KK, Vijayakumar M, Viveka K, et al. Fracturing of cutting balloon microsurgical blade inside coronary artery during angioplasty of tough restenotic lesion: A case report. Cathet Cardiovasc Intervent 2003;58:199–201.
5. Mehran R, Dangas G, Abizaid AS, et al. Angiographic patterns of in-stent restenosis: Classificaiton and implicaitons for long term outcome. Circulation 1999;100:1872–1878.
6. Montorsi P, Galli S, Fabbiocchi F, et al. Mechanism of cutting balloon angioplasty for in-stent restenosis: An intravascular ultrasound study. Cathet Cardiovasc Inervent 2002;56:166–173.

The Journal of Invasive Cardiology - ISSN: 1042-3931 - Volume 17 - Issue 1 (Jan 05) - January 2005 - Pages: 40 - 41



Novel Approaches to Managing Bradycardia during Coronary Rheolytic Thrombectomy

Special Supplement to the Journal of Invasive Cardiology


This special supplement was made possible through a grant from Possis Medical, Inc.
CLINICAL EXPERIENCE WITH A NEW HYBRID CORONARY WIRE

On Demand Web Archive
Non-Accredited


This activity is supported by an educational grant from Terumo Medical Corporation.
Pharmacoinvasive Management of Acute Coronary Syndrome: Incorporating the 2007 ACC/AHA Guidelines

Complimentary Accredited CME Program

This activity is supported by an educational grant from Sanofi-Aventis.
Varicose Veins: Causes, Symptoms, Diagnosis and Treatment of Chronic Venous Insufficiency

A Complimentary Accredited ON-DEMAND Webcast

This activity is supported by an educational grant from Diomed, Inc.


Create a Successful Vena Cava Filter Practice

Accredited CD

This activity is supported by an educational grant from Cook Incorporated and has been designed for Interventional Cardiologists, Vascular Surgeons, Fellows and Interventional Cardiovascular Nurses and Technologists.

Achieving Optimal Outcomes in Carotid Stenting: Lessons Learned from Recent Clinical Trials
Complimentary Accredited ON DEMAND Webcast

Topics
1. EVA-3S & Space-Bumps in the road
2. CAPTURE 3500-Lesion morphology & Predictors for Stroke
3. CAPTURE II vs. EXACT 1500-Does open or Closed Cell Stent design really matter?

This activity has been developed for Interventional Cardiologists, Vascular Surgeons, Interventional Radiologists, Neurologists, Interventional Nurses and Technologists with an interest in the diagnosis and treatment of peripheral artery disease.

Anticoagulation Techniques for Peripheral Vascular Interventions

Complimentary Accredited ON DEMAND Webcast

This activity has been developed for Interventional Cardiologists, Vascular Surgeons, Interventional Radiologists, Podiatric Physicians, Endovascular Allied Professionals, Endocrinologists, Wound Care Specialists, Directors of the Wound Care Clinic, and Primary Care Physicians, Pharmacists, Nurses and Technologists.


March 2007 Supplement
On-Demand Webcast
Archived Webcast

About HMP Communications

HMP Communications LLC (HMP) is the authoritative source for comprehensive information and education servicing healthcare professionals. HMP’s products include peer-reviewed and non-peer-reviewed medical journals, national tradeshows and conferences, online programs and customized clinical programs. HMP is a wholly owned subsidiary of HMP Communications Holdings LLC, which also owns the North American Center for Continuing Medical Education (NACCME). NACCME provides a wide array of accredited CME offerings with industry thought leaders participating in roundtable meetings, webcasts, symposia, conferences, seminars, podcasts and satellite programs. Discover more about HMP’s products and services at www.hmpcommunications.com. ©2008 HMP Communications


© 2008 HMP Communications | All Rights Reserved
83 General Warren Blvd, Suite 100 | Malvern, PA 19355 | 800.237.7285